1.1Open channel guides the instrument
A U-shaped polymer guide supports the instrument along its floor and sidewalls. The instrument slides along the channel, and the guide lifts sideways off the shaft after placement.
Generated automatically. Not reviewed by a person. Not legal advice or an opinion.
Questions about how individual parts of claim 1 might be changed. A question may describe something different from what the claim language recites, or a version of it; no view is offered on which.
A U-shaped polymer guide supports the instrument along its floor and sidewalls. The instrument slides along the channel, and the guide lifts sideways off the shaft after placement.
A low-profile balloon advances over the guidewire and inflates across the entry tract to enlarge it radially. After the balloon deflates and is withdrawn, a tapered sheath tip tracks over the same wire into the prepared tract.
The outside diameter of a thick-walled distal taper increases gradually from the wire opening to the full shaft diameter. After placement, the guidewire is withdrawn and a small-bore instrument passes through the fixed distal opening.
A single polymer extrusion gives both shaft segments the same material, bore diameter, and wall thickness. An external hub flange transfers connection loads to a skin-mounted holder.
A short ring embedded in the distal wall limits ovalization where instruments exit. The unreinforced proximal tube yields more under the same load, and a rounded polymer rim forms the tissue-contacting edge.
Further arrangements of the same subject matter, described here as of the publication date at the foot of this page.
A tubular guide sheath carries a removable tapered dilator through a sealed service port into a curved pipe branch. The sheath has a flexible thermoplastic tip and a stiffer thermoplastic base, so the tip deflects farther than the base under the same transverse force. An operator withdraws the dilator and advances a flexible inspection probe through the open lumen.
An inflatable tapered balloon on a catheter shaft replaces the solid dilator. The balloon mounts near the shaft tip, with its tapered section projecting beyond the sheath and its cylindrical section seated within the sheath tip. A soft polymer tip segment deforms farther under transverse load than the harder base segment. After placement, a syringe withdraws inflation fluid and the collapsed balloon passes back through the sheath lumen.
A slotted polymer sleeve replaces the homogeneous sheath wall, with an uncut base section and staggered circumferential slots in the tip section. A thin continuous elastomeric jacket spans the slots and seals the lumen from the outside. The polymer bridges between slots flex under transverse load, so the tip deflects farther than the uncut base. A smooth removable dilator slides through the bore and supports the slotted section during insertion.
A polymer sheath has a 2.0 mm bore, a 6 mm tip segment with a 0.15 mm wall, and a 0.30 mm base wall. The bore remains constant through the external transition, and a removable dilator with a 1.95 mm shaft slides through the lumen. The thinner tip wall deflects farther than the base wall when the same transverse force acts over the same axial contact width.
A hub bonded to the stiff base houses a slit elastomeric hemostasis valve and a side port between the valve and the sheath lumen. The dilator passes through the slit, which seals around the shaft and closes after withdrawal. The side port supplies flush fluid to the lumen, while a softer tip segment deflects farther than the base under the same transverse force.
A production fixture indexes an empty sheath onto opposed radiused anvils and tests the tip and base at separate axial positions. A motorized stage presses one anvil against each segment until a load cell reads the programmed transverse force, and a displacement encoder records the reduction in anvil spacing from first contact. The controller routes the sheath to dilator assembly when the tip compression exceeds the base compression by the specified minimum.
A thin fluoropolymer liner defines a circular bore along the sheath axis. An outer thermoplastic jacket captures the liner, and the bore diameter provides sliding clearance for the dilator.
A tapered polyethylene dilator extends beyond the sheath tip and contains a central guidewire bore. The dilator and sheath advance together over the wire, and the dilator withdraws through the proximal hub after placement.
Thermal bonding joins a lower-hardness distal polymer tube to a higher-hardness proximal tube with the same nominal bore and wall thickness. Parallel platens apply the same radial force to each segment, and the distal segment shows a larger decrease in diameter.
Passages reproduced word for word from two earlier-published U.S. patent documents, placed beside the language of claim 1. No characterization is offered. Each passage is marked A or B for the document it comes from; ¶ numbers are the paragraph numbers in the USPTO’s published full text of that document.
The tubular sheath member has a proximal end portion including a proximal end, a distal end portion terminating in a distal tip, and an internal sheath lumen extending through the proximal and distal end portions of the sheath member.B · US 2013/0165857 A1 · ¶ 4
The dilator is removably positioned within the sheath and includes a distal tapered end, a proximal end, and at least one blood detector.A · US 2012/0296275 A1 · ¶ 62
An empty cell means only that no passage was reproduced there.
Only Document A has a passage for 1.2. Only Document B has a passage for 1.1. Both are classified in CPC A61M, as is this patent.
Sentences reproduced word for word from these documents that state a problem, a purpose or benefit, a field of use, or an application or improvement.
In endoscopic and endovascular surgery, when access to hollow vessels is required via a conduit suitable for transporting or engaging therapeutic equipment and materials, introducer sheath-dilators are routinely used.A · US 2012/0296275 A1 · ¶ 3
Related claim parts: 1.1
In particular, introducer sheaths can be used by a clinician to access target areas of the patient's anatomy, e.g., the patient's cardiovascular system, and also allow for diagnostic and/or therapeutic payloads (such as ablation catheters) to be inserted into these target areas.B · US 2013/0165857 A1 · ¶ 3
Related claim parts: 1.2
Such materials may substantially increase the cost associated with the production of such devices and may provide undesirable geometric constraints that create obstacles in connection with delivering therapeutic equipment through the introducer sheath, particularly when such equipment or materials are non uniform in size and shape.A · US 2012/0296275 A1 · ¶ 3
In some embodiments of the present invention, a sheath dilator includes a blood detector, which may be used to calibrate the location of the sheath-dilator system with respect to an artery.A · US 2012/0296275 A1 · ¶ 4
Introducer sheaths often include hemostasis valves that permit introduction of the desired payloads but at the same time inhibit undesired fluid loss from the introducer sheath.B · US 2013/0165857 A1 · ¶ 3
patent.fyi, US 2026/0364313 A1, claim 1, first published 2026-11-23T01:04:04-05:00, https://patent.fyi/US20260364313/, SHA-256 fa092bfb92f94a334195c65c3fb66632b0e4632d888eeb9d9636a941267f9a40